PR Electronics PROFIBUS PA temperature transmitter PR 5450B

2026-03-09

5450_800x600px.jpg

PR5450B temperature transmitter

Environmental Conditions

Operating temperature -40°C to +80°C
Storage temperature -50°C to +85°C
Calibration temperature 23…25°C
Relative humidity < 99% RH (non-cond.)
Protection degree (encl./terminal) IP68 / IP00

Mechanical specifications

Dimensions Ø 44 x 21.45 mm
Center hole diameter Ø 6.35 mm / ¼ in
Weight approx. 50 g
Wire size 1 x 1.5 mm2 stranded wire
Stripping length 7 mm
Screw terminal torque 0.4 Nm
Vibration IEC 60068-2-6
2…25 Hz ±1.6 mm
25…100 Hz ±4 g

Common specifications

Supply

Supply voltage 9.0…30 VDC
Supply voltage in FISCO installations 9.0…17.5 VDC
Internal power dissipation < 352 mW

Isolation voltage

Isolation voltage, test / working 2.5 kVAC / 42 VAC

Response time

Response time < 400 ms
Programmable damping 0…60 s
Write protection Jumper
Warm-up time < 5 min.
Start-up time < 15 s
Programming PROFIBUS GSD, DD
Signal / noise ratio > 60 dB
Long-term stability, first year / 5 years ± 0.05% of reading / ± 0.10% of reading
Updating time, PROFIBUS < 100 ms
Signal dynamics, input 24 bit
Accuracy See manual for details
EMC immunity influence < ±0.1% of span
Extended EMC immunity: NAMUR NE21, A criterion, burst < ±1%

Input specifications

RTD input

RTD type Pt10…10000, Ni10…10000, Cu5…1000
Permissible cable resistance per wire ≤ 50 Ω
Effect of sensor cable resistance (3-/4-wire) < 0.002 Ω / Ω
Sensor current < 0.15 mA
Sensor error detection None, Shorted, Broken, Shorted or Broken

TC input

Thermocouple type B, E, J, K, L, N, R, S, T, U, W3, W5, LR
Cold junction compensation (CJC) Constant, internal or external via a Pt100 or Ni100 sensor
Sensor error detection None, Shorted, Broken, Shorted or Broken

Linear resistance input

Measurement range / min. range (span) 0 Ω…100 kΩ / 25 Ω
Permissible cable resistance per wire ≤ 50 Ω
Sensor current < 0.15 mA
Sensor error detection None, Broken

Potentiometer input

Potentiometer min….max. 10 Ω…100 kΩ
Measurement range / min. range (span) 0…100% / 10%
Permissible cable resistance per wire ≤ 50 Ω
Sensor current < 0.15 mA
Sensor error detection None, Shorted, Broken, Shorted or Broken

mV input

Measurement range -800…+800 mV (bipolar) -100…1700 mV
Min. measurement range (span) 2.5 mV
Input resistance 10 MΩ
Sensor error detection None, Broken

Output specifications

PROFIBUS PA connection

PROFIBUS PA protocol Profile A&B, ver. 4.0 (compatible with 3.02)
PROFIBUS PA protocol standard EN 50170 vol. 2
PROFIBUS PA address (at delivery) 126
PROFIBUS PA physical layer 2 x function blocks, 1 / 2 transducer block(s)
PROFIBUS PA termination 100 Ω +1 μF

国内销售