PR5725 Programmable frequency indicator
| Frequency range, f/I conversion function | 0.001 Hz to 50 kHz |
| Low cut-off frequency | 0.0009 Hz (default value) |
| Max. frequency, with input filter ON | 50 Hz |
| Time range, period time function | 999.9 s to 20 μs |
| Low cut off period time (time-out) | 1111 s |
| Min. period time with input filter ON | 20 ms |
| Input types | NAMUR acc. to EN 60947-5-6 Tacho NPN / PNP TTL S0 acc. to DIN 43864 Special voltage Special current |
Display |
|
| Display readout | -1999…9999 (4 digits) |
| Decimal point | Programmable |
| Digit height | 13.8 mm |
| Display updating. | 2.2 times / s |
| Display response time, programmable | 0.0…60.0 s |
| Input outside input range is indicated by | Explanatory text |
Current output |
|
| Programmable signal ranges | 0…20/4…20/20…0/20…4 mA |
| Load (@ current output) | ≤ 800 Ω |
| Load stability | ≤ 0.01% of span / 100 Ω |
| Current limit | ≤ 28 mA |
| Sensor error indication | 0 / 3.5 / 23 mA / none |
| Output limitation, on 4…20 and 20…4 mA signals | 3.8…20.5 mA |
| Output limitation, on 0…20 and 20…0 mA signals | 0…20.5 mA |
Relay output |
|
| Relay functions | Setpoint |
| Hysteresis, in % / display counts | 0…100% / 0…9999 |
| ON and OFF delay | 0…3600 s |
| Power On delay | 0…60 s |
| Sensor error reaction | Break / Make / Hold |
| Max. voltage | 250 VAC / VDC |
| Max. current | 2 A |
| Max. AC power | 500 VA |
| Max. DC current, resistive load > 30 VDC | See manual for details |